- Title
- 3D dose reconstruction: banding artefacts in cine mode EPID images during VMAT delivery
- Creator
- Woodruff, H. C.; Greer, P. B.
- Relation
- 7th International Conference on 3D Radiation Dosimetry (IC3DDose). Proccedings of the 7th International Conference on 3D Radiation Dosimetry (IC3DDose) (Sydney, AUSTRALIA 4-8 November, 2012)
- Publisher Link
- http://dx.doi.org/10.1088/1742-6596/444/1/012042
- Publisher
- Institute of Physics Publishing
- Resource Type
- conference paper
- Date
- 2013
- Description
- Cine (continuous) mode images obtained during VMAT delivery are heavily degraded by banding artefacts. We have developed a method to reconstruct the pulse sequence (and hence dose deposited) from open field images. For clinical VMAT fields we have devised a frame averaging strategy that greatly improves image quality and dosimetric information for three-dimensional dose reconstruction.
- Subject
- cine mode images; VMAT delivery; banding artefacts; open field images
- Identifier
- http://hdl.handle.net/1959.13/1062558
- Identifier
- uon:17106
- Identifier
- ISBN:9781627486767
- Language
- eng
- Full Text
- Reviewed
- Hits: 2688
- Visitors: 2899
- Downloads: 252
Thumbnail | File | Description | Size | Format | |||
---|---|---|---|---|---|---|---|
View Details Download | ATTACHMENT01 | Publisher version (open access) | 709 KB | Adobe Acrobat PDF | View Details Download |